Recalls for H&H Medical Corporation

Emergency Cricothyrotomy Kit By H&H Medical Corporation: Recall - Defective Cuff Balloon On Endotracheal Airway
10/04/2013 Medical Recall

Defect due to a reduction in package density, a higher than average dose of gamma sterilization, and the occasional slippage of a protective silicon sleeve during shipping.

  • Company:
    H&H Medical Corporation
H&H Medical Corporation Issues Nationwide Recall of Emergency Cricothyrotomy Kit
10/04/2013 Food Recall

On August 27, 2013, H&H Medical Corporation initiated a nationwide recall of 6,619 units of the H&H Emergency Cricothyrotomy Kit. The product has been found to show the potential for a defective cuff balloon on the provided endotracheal airway.

  • Company:
    H&H Medical Corporation